Water Treatment Questions and Answers
100% Pass
2 reasons for water Tx ✔✔1- Pt safety
2- Prevent equipment damage
Aluminum in water used for dialysis causes ✔✔Anemia
Bone disease
nausea
vomiting
Chlorine in water
...
Water Treatment Questions and Answers
100% Pass
2 reasons for water Tx ✔✔1- Pt safety
2- Prevent equipment damage
Aluminum in water used for dialysis causes ✔✔Anemia
Bone disease
nausea
vomiting
Chlorine in water used for dialysis causes ✔✔hemolysis
Excess calcium and sodium in water used for dialysis causes ✔✔hypertension
Excess calcium and magnesium in water used for dialysis causes ✔✔muscle weakness
Water/Dialysate culture results- acceptable level ✔✔<50 cfu/ml
Water/Dialysate culture results- Action level ✔✔50-199 cfu/ml
Water/Dialysate culture results- Unacceptable level ✔✔>/= 200 cfu/ml
Water/Dialysate endotoxin testing results - acceptable level ✔✔<1 EU/ml
Water/Dialysate endotoxin testing results- action level ✔✔1-2 EU/ml
Water/Dialysate endotoxin testing results- unacceptable level ✔✔>2 Eu/ml
What removes chlorine/chloramines ✔✔carbon tanks
Chlorine/Chloramines testing- When ✔✔After RO runs for 15 mins before 1st shift and every 4
hrs
If tested prior to 15 mins can produce false neg. result
Chlorine/Chloramines testing- Where ✔✔after primary carbon tank
Chlorine/Chloramines testing- acceptable result ✔✔= 0.1 mg/L
Chlorine/Chloramines testing- Action if primary tank too high ✔✔repeat test
if still high test secondary tank
Chlorine/Chloramines testing- F/U secondary tank with limits ✔✔notify biomed and FA
Test every 30 mins
Chlorine/Chloramines testing- F/U secondary tank too high ✔✔Stop dialysis
notify biomed and FA
Hardness Testing- What is removed ✔✔Calcium and Magnesium
Hardness Testing- Why is it important ✔✔RO protection
Hardness Testing- When do we test ✔✔At the end of the day
What does the RO remove ✔✔organic and inorganic materials bacteria and endotoxins
What are the 2 concerns w/using DI tanks ✔✔1- exhaust quickly
2- highly dangerous when exhausted
Response to Final Water quality alarm ✔✔1- Place all machines in bypass
2- Notify biomed, FA, RN, MD
3- If not restored, terminate treatments
[Show More]