Lab 5: Flow Rate Sensor Calibration
Sangyun Go
[email protected]
Oct 1st, 2015
Lab section 001
1Flow Rate Sensor Calibration
List of Figures
Figure 1: Turbine Flow Meter (GPM) vs. Rotameter (GPM ) …….…….…………….3
...
Lab 5: Flow Rate Sensor Calibration
Sangyun Go
[email protected]
Oct 1st, 2015
Lab section 001
1Flow Rate Sensor Calibration
List of Figures
Figure 1: Turbine Flow Meter (GPM) vs. Rotameter (GPM ) …….…….…………….3
Figure 2: Turbine Flow Meter (GPM) vs. Vortex Shedding Meter (GPM) …….…….…………….4
List of Tables
Table 1: Measured values for flow sensor calibration ….……….……….……..............................1
Table 2: Interpolation Data For Turbine Flow Meter ….……….……….……..............................2
Table 3: Flow Values for Sensor Calibration ….……….……….……..............................3
Table 4: Rotameter Values for Calculations ….……….……….……..............................5
Table 5: Vortex Shedding Meter Values for Calculations ….……….……….……..............................5
Table 6: Uncertainty Values ….……….……….……..............................6
Table 7: Calibration Curve Equations ….……….……….……..............................7
23Flow Rate Sensor Calibration
Abstract/Objectives
Sensors always come with an uncertainty value that is rated by the manufacturer. This lab compares this
rated accuracy with a calculated accuracy based off measurements that were made. Each set of data is
plotted and a linear fit of the data was used to calculate the sensors calibration function. With this function,
the measurements can be corrected during data analysis. Once the lab is completed, the results were
tabulated and a comparison was made between the uncertainty values.
Procedure
In preparation for lab, the manual for lab three should be read out of the lab manual “Measurement and
Instrumentation Laboratory Manual” by John F. Maddox and Jordan C. Roberts. In the lab, a rotameter,
turbine flow meter, and vortex shedding meter were setup along a piping system in series. Water was drawn
from a container by a centrifugal pump at a steady rate along the system. Using a valve, the flow rate was
adjusted to certain percentages of maximum flow rate. These percentages were read on the rotameter. With
each percentage, data was collected from each sensor. These results are shown in Table 1
[Show More]